Attention
1.Make sure there is no ash on the connector and check the screw thread,then install the inhalator.
2.Make sure the flow meter is vertical to the floor and flow regulating valve is closed,then open the valve of oxygen cylinder slowly,regulate the flow after the high pressure gauge shows the pressure of the oxygen cylinder.
3.No drastic shock,far away with the corrosive substance,and no oil stain.
4.The set screw of the safety valve can't be rotated at random,non-professional people are forbidden to dismount.
Type | Gas Type | Input Pressure | Output Pressure | Control Pressure of Safety Valve | Flowmeter Control Range | Material | Connecting thread |
DY-C | Oxygen | 150bar | 2bar-3bar | 3.5bar±0.5bar | 1L/min-10L/15L/min | Brass | Female G5/8 Male G5/8 G3/4 |
DY-C1 | |||||||
DY-C2 | |||||||
DY-C3 | |||||||
DY-C4 | |||||||
DY-C5 |